Comparison of the accuracy of different handheld-type scanners in three-dimensional facial image recognition
Yumi TsuchidaMaho ShiozawaKazuyuki HandaHidekazu Takahashi Hiroki Nikawa
著者情報
ジャーナル オープンアクセス

2023 年 67 巻 2 号 p. 222-230

詳細
抄録

Purpose: Handheld-type scanners are widely used in clinical practice. This study examined the accuracy of handheld-type scanners using plaster statues to assess their performance in facial recognition.

Methods: Twelve 4-mm zirconia balls as measuring points were attached to the facial portions of three types of plaster statue. Six digital facial images of each plaster statue were obtained using one of the following five handheld-type scanners: Artec Eva, Artec Spider, Bellus 3D FaceApp, SNAP, and Vectra H1. Four-millimeter spherical objects were manually placed at the measurement points on the scanned data generated using computer-aided design software and coordinate positions were measured using a contact-type high-resolution three-dimensional measurement device. Consequently, the discrepancy between the distance measured using the contact-type device and that measured using the handheld-type scanner was calculated. The scanning time, processing time, and deviation of the distance between the measuring points were analyzed using two-way analysis of variance and t-test with Bonferroni correction.

Results: The scanning and processing times ranged from 15.2 to 42.2 s and 20.7 to 234.2 s, respectively. Overall, 97% of all measured distances by Spider were within ±1.00% deviation; 79%, Vectra; 73%, Eva; 70%, Bellus; and 42%, SNAP.

Conclusions: The performance of handheld-type scanners using plaster statues varied among the different scanners. The scanning time of Eva and the processing time of Bellus were significantly shorter than those of other scanners. Furthermore, Spider exhibited the best accuracy, followed by Eva, Vectra, Bellus, and SNAP.

著者関連情報
© 2023 Japan Prosthodontic Society

This is an open-access article distributed under the terms of Creative Commons Attribution-NonCommercial License 4.0 (CC BYNC 4.0), which allows users to distribute and copy the material in any format as long as credit is given to the Japan Prosthodontic Society. It should be noted however, that the material cannot be used for commercial purposes.