Exploring the bio-mechanical behavior of PEEK and CFR-PEEK materials for dental implant applications using finite element analysis
Kandula Uday Kumar ReddyAqshat SethAmol VuppuluriPiyush Chandra VermaSuresh Kumar Reddy NaralaPolavarapu Jayakrishna BabuPrabakaran Saravanan
著者情報
ジャーナル オープンアクセス

2024 年 69 巻 1 号 p. 41-48

詳細
抄録

Purpose: This study explored the bio-mechanical properties of polyether ether ketone (PEEK) and carbon fiber reinforced-PEEK (CFR-PEEK) as potential alternatives to traditional dental implant materials, such as titanium (Ti) and zirconia (ZrO2). Conventional implant materials often exhibit stress shielding leading to peri-implant bone loss and implant failure.

Study selection: Finite element analysis using a three-dimensional computer-aided-design (3D CAD) model of the jawbone with various implant materials (titanium, zirconia, PEEK, and CFR-PEEK) incorporated was implemented to assess the effectiveness of PEEK and CFR-PEEK. Two loading conditions (50 and 100 N) were applied in centric (case-1) and eccentric (case-2) to mimic the oral loading conditions.

Results: Titanium and zirconia implants were found to exhibit higher levels of stress shielding and therefore pose greater risks of bone loss and implant failure. Conversely, CFR-PEEK implants demonstrated more-uniform stress distributions that reduce the likelihood of stress shielding compared to their conventional counterparts; consequently, CFR-PEEK implants are particularly suitable for load-bearing applications. Furthermore, maximum strain values on PEEK-implanted cortical bone reached a state of adaptation, referred to as the “lazy zone” in which bone growth and bone loss rates are equal, indicating PEEK’s potential for a long-term implant utilization.

Conclusions: PEEK and CFR-PEEK implants are promising alternatives to conventional dental implants because they provide stress shielding and promote bone health. Improved stress distribution enhances long-term success and durability while mitigating complications, and highlights their applicability to dental implant procedures.

著者関連情報
© 2025 Japan Prosthodontic Society

This is an open-access article distributed under the terms of Creative Commons Attribution-NonCommercial License 4.0 (CC BYNC 4.0), which allows users to distribute and copy the material in any format as long as credit is given to the Japan Prosthodontic Society. It should be noted however, that the material cannot be used for commercial purposes.