Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Experimental studies on platelet function and circulatory status during extracorporeal circulation (Comparison Hollow-fiber membrane Oxygenator and bubble Oxygenators)
Harunori TAKEDAHiroyoshi MATSUKURAToshiaki KAWAKAMITatsuzo TANABE
Author information
JOURNAL FREE ACCESS

1983 Volume 12 Issue 2 Pages 441-444

Details
Abstract
In 16 mongel dogs, 2 hour total cardiopulmonary bypasses were carried out with Hollow-fiber oxygenators and bubble oxygenators for evaluation of platelet function and circulatory status. Adequet oxygenation and carbon deoxide elimination were observed with Hollow fiber oxygenators in spite of low ratio of oxygen flow to blood flow.
Plasma free hemoglobin level was low throughout perfusion compared to bubble oxygenators with Hollow fiber oxygenators. Platelet kinetics were elevated during perfusion. It seems to suggest that platelets in active stage are brought back into circulatory blood by detachment or deaggultination.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback
Top