人工臓器
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Development of an Ascending Aortic Valved Conduit
Andy CAMPBELLJoseph SAUTERHitoshi KOYANAGI
著者情報
ジャーナル フリー

1992 年 21 巻 2 号 p. 625-634

詳細
抄録
A long tapered section is less convenient for coronary anastomosis, but conduit manufacturers argue that it optimizes blood flow. In this study, flow through anatomical and conduit models was observed and flow through model coronary arteries was measured. Comparison indicated that there is no advantage to a long tapered section with regard to flow. A minimized tapered section is therefore optimal.
著者関連情報
© The Japanese Society for Artificial Organs
前の記事 次の記事
feedback
Top