Geometrical fault in basic design of hollow fiber type dialyzer
Iwakazu KANEKOTetsuzo AGISHIKatsuhito YAMAGATAToshihisa HOSHINOKazuo ERAKazuo OTA
Author information
JOURNAL FREE ACCESS

1978 Volume 7 Issue 5 Pages 645-647

Details
Abstract
Some contrivances are made on hollow fiber type dialyzer in an attempt to increace dialysis performance.
Of these, gadgets in a dialysate compartmente such as construction of baffle(s)or a long dialysate channel are evidenced to present a dangerous problem under particular workingcondition. Flow resistance is higer in a dialysate compartment than in a blood compartment. As a result, ultrafiltration directs from a dialysate compartment to ward a blood compartment.
If there are any broken fibers, nonsterile dialysate directly flows into blood.