Japanese Journal of Clinical Chemistry
Online ISSN : 2187-4077
Print ISSN : 0370-5633
ISSN-L : 0370-5633
Continuous Flow Injection Analysis of Total Bile Acid in Serum
TATSUMORI TAKEDASHIGERU YOSHIDAKAYOKO ODASHINGO HIROSE
Author information
JOURNAL FREE ACCESS

1984 Volume 13 Issue 3 Pages 134-139

Details
Abstract
A highly sensitive fluorescence assay for total bile acid in serum using a flow injection system is described.
The applrltus consists of six-port valve, pretreatment column (2×20mm) packed serumout-25 (Sekisui chemical co.), immobilized enzyme column (3α-hydroxysteroid dehydrogenase, 3α-HSD, 2×30mm), injector, peristaltic pump and fluorescence detector.
This system enables to detect total bile acid from serum injection directly without sample pretreltment.The new device is especially useful for clinical studies where total bile acid has to be determined from serum.Moreover, the method was more sensitive and simple than conventional method.(cv=1.99%, n=10).