臨床化学
Online ISSN : 2187-4077
Print ISSN : 0370-5633
ISSN-L : 0370-5633
Stream Switching法によるCPK-BBの活性測定とその評価
武田 立守吉田 滋織田 佳代子広瀬 信吾
著者情報
ジャーナル フリー

1984 年 13 巻 5 号 p. 258-263

詳細
抄録
We investigated CPK.BB activity in serum of various individuals at Ueno Civic Hospital.CPK-BB activity was determined by Stream Switching System containing micro colum packed DEAE-sepharose CL 6B, six-port valve, injector and fluorescence detector (Ex 360nm, Em 460nm).The system was useful for the correction of background caused serum and reagents.
The CPK-BB activities of healthy individuals were0.86mU/ml (CPK-BB/total CPK=0.0098) for female (n=35) and 0.92 mU/ml (CPK-BB/total CPK=0.0060) for male (n=35) respectively.We found abnormaiities of CPK-BB activity and relative activity (CPK-BB/total CPK) in patient's serum of renal failure, prostatic cancer, gastric cancer and so on.Especially, in the case of renal failure, CPK-BB activity was determined sensitively without the effects of artifact like vitamines and atbumin in serum.
著者関連情報
© 日本臨床化学会
前の記事 次の記事
feedback
Top