臨床化学
Online ISSN : 2187-4077
Print ISSN : 0370-5633
ISSN-L : 0370-5633
抱合体加水分解不要な尿中17-ケトステロイド測定法の考案
野崎 修大畠 恒子大場 康寛森山 弘之加藤 芳男
著者情報
ジャーナル フリー

1989 年 18 巻 2 号 p. 49-54

詳細
抄録
We developed a method for measuring urinary 17 ketosteroids (17 KS) without prior hydrolysis of its conjugated forms. Both free and conjugated forms of 17 KS were successfully extracted by passing the samle through a BSA-ODS minicolumn (TSK), eluting with 80% methanol and drying. After the Zimmermann reaction, the Zimmermann chromogen was extracted with 1-butanol. Our proposed method gave rapid results for urinary 17 KS and showed good reproducibility (C. V.: 2.41-8.75%), recovery (about 100%) and correlation with the conventional method for determining 17 KS (r=0.881, a regression line, y=0.565x+4.52, n=25).
著者関連情報
© 日本臨床化学会
前の記事 次の記事
feedback
Top