臨床化学
Online ISSN : 2187-4077
Print ISSN : 0370-5633
ISSN-L : 0370-5633
High-Performance Liquid Chromatography of Estrogen 3-Sulfates with Electrochemical Detection Using an Immobilized Enzyme Reactor
Toshifumi NiwaTomoyoshi KoshiyamaShunji IshiwataJunichi GotoToshio Nambara
著者情報
ジャーナル フリー

1991 年 20 巻 4 号 p. 242-246

詳細
抄録

A simple and sensitive method for the determination of estrogen 3-sulfates by highperformance liquid chromatography with electrochemical detection is described. The sulfates were separated on a reversed-phase column and subsequently hydrolyzed with an immobilized sulfatase reactor in a continuous flow mode. Estrogens liberated were then detected with a coulometric detector. A linear response was observed for four sulfates, estrone sulfate, 16 a -hydroxyestrone 3-sulfate, estradiol 3-sulfate and estriol 3-sulfate in the range 25-1000 ng as an injected amount. The detection limits (S/N=10) were 20 ng for estriol 3-sulfate and 16 a -hydroxyestrone 3-sulfate, and 50 ng for estrone sulfate and estradiol 3-sulfate, respectively.

著者関連情報
© Japan Society of Clinical Chemistry
前の記事 次の記事
feedback
Top