バイオエンジニアリング講演会講演論文集
Online ISSN : 2424-2829
セッションID: 108
会議情報
108 In-vitro可視化実験による下大静脈フィルターの血栓捕獲性能評価(OS1-2:循環器系・血流の計測とメカニクス(2),OS1:循環器系・血流の計測とメカニクス)
Jin-long LIUYing-zheng LIUZhao-Min CAO銭 逸梅津 光生Zhi-Min FEI
著者情報
会議録・要旨集 フリー

詳細
抄録
Flow performance of three VCFs placed in vein was evaluated in an in-vitro experimental test apparatus. Toward this end, a low-speed recirculating water channel was established. Three VCFs (Greenfield filter, TrapEase filter and ALN filter) were chosen for an extensive comparison at Re=500〜700. Nine types of in-house made cylindrical materials with different diameter and length were used to model the thrombus. Efficiency of trapping the different sized thrombus was obtained; pressure drop of the flow through VCFs was measured. A detailed view of the local complex flow over the VCFS with trapped thrombus was revealed by flow visualization. The in-vitro experimental data are of crucial significance to the clinic purpose.
著者関連情報
© 2008 一般社団法人 日本機械学会