p. 25-26
Achieving high control accuracy is a key issue for a surgical intravascular microrobot. This paper discusses the identification and optimization of the gains of a PI controller designed for the control of electromagnetic field for operating a 1 mm diameter microrobot. The microrobot was successfully controlled with a higher accuracy after tuning the control parameters.