S021011 拍動流下における連続流型補助人工心臓の耐久性試験方法に対する検討([S02101]人工臓器におけるバイオメカニクス(1))
西田 正浩小阪 亮丸山 修山根 隆志大久保 剛星 英男中島 祥吾長田 俊幸妙中 義之
著者情報
会議録・要旨集 フリー

詳細
抄録
Durability test mock circulations for a newly developed axial-flow ventricular assist device with hydrodynamic bearings have been developed to start a 2-year durability test under pulsatile flow condition. The mock circulation with a diaphragm pump was adjusted to achieve a pulsatile flow rate of 0 L/min in the diastolic phase and 9.5 L/min in the systolic phase filled with saline added with glycerin as a blood analog fluid. The 268-days pre-test for the axial-flow ventricular assist pump was conducted with the mock circulation. As a result, although the test has maintained the pump flow rate, head, fluid temperature and density, pulse rate of the diaphragm pump, a slight Fe component was solute into the working fluid. Therefore, the mechanical heart valve was substituted for the check valve for the mock circulation.