1106 有限要素法と粒子画像流速計を用いたダイアライザ内の流れ解析(S14-2 制御と情報・生体への応用(2),S14 制御と情報・生体への応用)
石田 等佐藤 敏夫辻 毅一川島 徳道阿岸 鉄三
著者情報
会議録・要旨集 フリー

詳細
抄録
Visualization and numerical analysis of the flow in a hollow fiber type hemodialyzer were conducted to investigate the channeling phenomenon in which the blood flow deviates from normal behavior. Dialysate flow is perturbed by baffle to make the function of dialyzer ineffective. Dialysate flow was visualized with a Particle Image Velocimeter (PIV) after the baffle structure was mechanically changed. The dialysate and blood flow were numerically analyzed using a Finite Element Method (FEM). As a result, we could propose an effective configuration of dialyzer that has advantages in both traditional and improved types.
著者関連情報
© 2005 一般社団法人日本機械学会