マイクロ・ナノ工学シンポジウム
Online ISSN : 2432-9495
セッションID: 5PM3-PMN-051
会議情報
5PM3-PMN-051 漏れ電界を用いた液体分析用デュアル水晶センサの開発(マイクロセンサとナノ計測,ポスターセッション)
今井 寛明武石 大一寒川 雅之安部 隆
著者情報
会議録・要旨集 フリー

詳細
抄録
QCM (Quartz crystal microbalance) is used as viscoelastic sensor not only as mass sensor. If it can measure the electric properties, it is expected to use for various applications. In this paper, a dual quartz sensor using a leakage electric field has been developed for analyzing the physicochemical properties of liquids. This sensor can measure the mechanical properties and the electrical properties simultaneously by integrating two types of ACR (Antiparallel coupled resonator) sensors on a same substrate. One is the sensor for measuring the viscoelasticity, another is the sensor for measuring the conductivity and dielectric constant. By changing the electrode portion on the sensing surface of ACR, ACR can evaluate the electrical properties of liquid using a leakage electric field. And this sensor can improve its stability in the air by adjusting the electrode area on the sensing surface.
著者関連情報
© 2013 一般社団法人 日本機械学会
前の記事 次の記事
feedback
Top