This study creates a blood vessel model for a catheter surgery simulator. In previous study was succeeded in producing a blood vessel model applicable to a surgical simulator using PVA hydrogel. On the other hand, the problem with the blood vessel model is that the reproducibility of the mechanical properties of blood vessels, which is important for the actual catheter operation sensation, is low. Therefore, in this study, creates a blood vessel model with a two-layer structure that reproduces the mechanical properties of the blood vessel wall and adipose tissue. In the proposed blood vessel model, it is shown that the mechanical properties can be reproduced by changing the compounding ratio of powdered PVA and the mass ratio of water and DMSO as a solvent. Furthermore, this study proposes a method for producing a two-layer structure and show that it is possible to produce a blood vessel model with a two-layer structure.