ロボティクス・メカトロニクス講演会講演概要集
Online ISSN : 2424-3124
セッションID: 1P2-W04
会議情報
1P2-W04 巻数増加と抵抗値低減を両立可能な折り曲げ式MRI用マイクロコイル
岩永 麗樹土肥 徹次
著者情報
会議録・要旨集 フリー

詳細
抄録
This paper reports on a folding micro coil as a receiver for MRI (Magnetic Resonance Imaging). A folding micro coil was fabricated by aligning uniformly circular plates along the coil axis by using MEMS (Micro Electro Mechanical Systems) technology. Since line width depends only on the direction of coil diameter, the line width doesn't decrease with increasing coil turns. Therefore we can fabricate the micro coil which has compatibility between increasing coil turns and decreasing resistance. The diameter of the coil and the length of the coil were 6 mm and 12 mm. The number of coil turns were 4, 6, and 8. The resistance of turn 4, turn 6, and turn 8 coils were 1.24 Ω, 1.54 Ω, and 1.69 Ω, respectively. Since a folding micro coil has low resistance, we acquired MR images with high SNR.
著者関連情報
© 2015 一般社団法人 日本機械学会
前の記事 次の記事
feedback
Top