日本口腔インプラント学会誌
Online ISSN : 2187-9117
Print ISSN : 0914-6695
ISSN-L : 0914-6695
原著
インプラントにおける磁性アタッチメントの応用:MRIへの影響の検討
山城 光明田中 譲治金田 隆星野 和正小澤 幸重
著者情報
ジャーナル フリー

2001 年 14 巻 2 号 p. 189-196

詳細
抄録
Recently, magnetic attachments for implants have been reported to be useful. However, keepers of magnetic attachments have produced metal artifacts on magnetic resonance (MR) imaging in the head and neck regions. The purpose of this study was to evaluate metal artifacts on MR imaging using Magnetic Attachments of a Cap Shape (MACS) removable keepers, in a phantom study and in clinical cases.
The following results were obtained:
1. In the phantom study, the dimensions of metal artifacts were 11.3 centimeters with a keeper,and almost none without a keeper.
2. In clinical case, diagnostic imaging in the head and neck regions was impossible with keepers, and possible without keepers. These results suggested that MACS is useful for MR imaging in the oral region, because it has a removable keeper.
著者関連情報
© 2001 公益社団法人日本口腔インプラント学会