水質汚濁研究
Print ISSN : 0387-2025
純水に対する溶存酸素濃度の測定
主として溶解度について
斉藤 紘史内山 英樹田中 良行
著者情報
ジャーナル フリー

1983 年 6 巻 4 号 p. 237-244

詳細
抄録

This report describes the experimental procedure and the measurement of oxygen solubility (dissolved oxygen concentration with air) in pure water, which is used for the calibration of dissolved oxygen meters. The solubility in the region of 0-40°C C was measured by means of modified Winkler titrimetric technique in which the iodine volatilization was controlled.
Experimental result showed that the solubility c5 (mg/l) at temperature t (°C) was expressed with following polynominal equation and its accuracy was better than ±0.4 %.
c5 = 14.536- 0.4049t+ 8.520×10-3t2-1.123×10-4t3 + 6.34×10-7t4.
The influences, of the iodine volatilization, the atmospheric pressure, and dissolved oxygen in reagents on the measured value were discussed and the result was compared with the previous works.

著者関連情報
© 社団法人日本水環境学会
前の記事 次の記事
feedback
Top