Comparison of Extracapillary and Endocapillary Blood Flow Oxygenators for Open Heart Surgery in Dogs: Efficiency of Gas Exchange and Platelet Conservation
Katsuichiro HOSHIRyou TANAKAAkira SHIBAZAKIYukiko NAGASHIMAHidehiro HIRAORyosuke NAMIKIKazuaki TAKASHIMAYasuharu NOISHIKIYoshihisa YAMANE
著者情報
ジャーナル フリー

2003 年 65 巻 3 号 p. 357-361

詳細
抄録

The goal of the current study was to compare the efficiency of gas exchange and platelet conservation of a new extracapillary blood flow oxygenator versus an endocapillary blood flow oxygenator during open heart surgery with extracorporeal circulation in dogs. Dilation and remodeling of the right ventricular outflow tract of dogs was performed using a patch graft technique to simulate pulmonary stenosis. Sequential pre- and post-operative blood analysis revealed that gas exchange efficiency and platelet conservation was significantly greater with the extracapillary blood flow oxygenator than with the endocapillary blood flow oxygenator. However, the priming volume of the extracapillary blood flow oxygenator was significantly greater, leading to hemodilution. We conclude that while the extracapillary blood flow oxygenator provided benefits in terms of gas exchange and platelet conservation, development of a smaller extracapillary blood flow type oxygenator to reduce hemodilution effects would be beneficial.

著者関連情報
© 2003 by the Japanese Society of Veterinary Science

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.ja