The Interfacial Strength in Sputtering-Hydroxyapatite-Coating Implants with Arc-Deposited Surface
Takahiro SENOYasuharu IZUMISAWAIkuya NISHIMURASeiya MAEHARATokiko YOKOYAMAKazuto YAMASHITATadao KOTANI
Author information
JOURNAL FREE ACCESS

2003 Volume 65 Issue 3 Pages 419-422

Details
Abstract

Titanium columns (Ti-6A1-4V) were treated with arc-deposition to roughen the surface enough to anchor the bone, and then coated with hydroxyapatite (HA) at a thickness of 5 μm by the sputtering technique. Columns were implanted into dog femurs, and fixation of columns to bone due to bone-ingrowth was assessed histologically and with the push-out test. The HA-coated columns were inserted in the shafts of the right femurs of 2 dogs. As a control, columns that were only arc-deposited (non-coated columns) were inserted into the left femurs. The interfacial strength was higher for the HA-coated columns than for the non-coated columns. Coating a rough surface with an HA layer using a sputtering technique reinforces interfacial strength between bone and implants.

Content from these authors
© 2003 by the Japanese Society of Veterinary Science

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.ja