Journal of Veterinary Medical Science
Online ISSN : 1347-7439
Print ISSN : 0916-7250
ISSN-L : 0916-7250
FULL PAPER
Minimally Invasive Determination of Cardiac Output by Transthoracic Bioimpedance, Partial Carbon Dioxide Rebreathing, and Transesophageal Doppler Echocardiography in Beagle Dogs
Kazuto YAMASHITAYukie UEYAMAKenjiro MIYOSHIRitsuyo IGARASHITokiko KUSHIROMohammed A. UMARWilliam W. MUIR
Author information
JOURNAL FREE ACCESS

2007 Volume 69 Issue 1 Pages 43-47

Details
Abstract

Minimally invasive cardiac output was determined using transthoracic bioimpedance (BICO), partial carbon dioxide rebreathing (NICO) and transesophageal Doppler echocardiography (TEECO) and compared to thermodilution (TDCO) in 6 beagle dogs. The dogs were 2 years old, weigh between 9.1-13.0 kg and were anesthetized with nitrous oxide-oxygen-sevoflurane. All dogs were administered a neuromuscular blocking drug and artificially ventilated during anesthesia. Thirty paired measurements of TDCO and each non-invasive method were collected during low, intermediate, and high values of cardiac output achieved by varying the depth of anesthesia and the administration of dobutamine. Cardiac output values ranged from 1.10-2.50 L/min for BICO compared to 0.81-4.88 L/min for TDCO; 0.70-2.60 L/min for NICO compared to 0.89-4.45 L/min for TDCO; and 0.59-4.37 L/min for TEECO compared to 0.57-4.15 L/min for TDCO. The limits of agreement and percentage error were -0.58 ± 1.56 L/min and ± 75.4% for BICO, -1.04 ± 1.08 L/min and ± 56.0% for NICO, and 0.03 ± 0.26 L/min and ± 12.3% for TEECO compared to TDCO. In conclusion, TEECO provided the best agreement to TDCO in sevoflurane anesthetized beagle dogs.

Content from these authors
© 2007 by the Japanese Society of Veterinary Science

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.ja
Previous article Next article
feedback
Top