In Vitro and In Vivo Comparison of Applanation Tonometry and Rebound Tonometry in Dogs
Noriyuki NAGATAMasashi YUKITakashi HASEGAWA
著者情報
ジャーナル フリー

2011 年 73 巻 12 号 p. 1585-1589

詳細
抄録

Intraocular pressure (IOP) evaluated by applanation tonometry via TONO-PEN XL (TP), and rebound tonometry via TonoVet (TV) were compared in enucleated canine eyes with varied pressure of the anterior chamber (AC) and in clinical cases. TV measured IOP values were lower than IOP measurements of TP in the enucleated eyes with 5-10 mmHg of AC (P<0.0001), though there was no significant difference in IOP values obtained with TP and TV on the pressure ranges of 15-20 mmHg. However, TP detected IOP values were lower than IOP measurements of TV in the eyes with over 25 mmHg of AC (P<0.0001). The results of clinical cases were similar to the enucleated eye model. There was no significant difference in IOP values obtained from TP and TV in dogs with normotensive eyes. IOP measurements of TP were lower than those of TV in glaucomatous eyes (P<0.0001). TV was a reliable tonometer for measurement of IOP in hypertensive eyes, whereas it was less accurate than TP in hypotensive eyes. The characteristics of TP and TV should be considered in the evaluation of IOP in practice.