Journal of Veterinary Medical Science
Online ISSN : 1347-7439
Print ISSN : 0916-7250
ISSN-L : 0916-7250
Physiology
Commercial human kits’ applicability for the determination of biochemical parameters in sheep plasma
Valeria PASCIUElena BARALLAMaria NIEDDUSara SUCCUCristian PORCUGiovanni G. LEONIPietro SECHIGiovanni C. BOMBOIFiammetta BERLINGUER
Author information
JOURNAL FREE ACCESS

2018 Volume 81 Issue 2 Pages 294-297

Details
Abstract

Currently, several commercially available biochemical kits are validated for their use in human but not in animals. The purpose of this work is to demonstrate the applicability of human kits for alanine-aminotransferase, aspartato-aminotransferase, albumin, total protein, total cholesterol, and triglycerides in ovine plasma. Assays were validated according to international guidelines and stability was explored. Accuracy values were between 67 and 100%, and intra and interday precisions (%RSD) were <15% for all studied parameters. These results confirm the suitability of the studied human kits for their use in ovine plasma and they were used in plasma collected from pregnant ewes.