Journal of Veterinary Medical Science
Online ISSN : 1347-7439
Print ISSN : 0916-7250
ISSN-L : 0916-7250

This article has now been updated. Please use the final version.

Detecting endotoxin activity in bovine serum using an automated testing system
Kazuyuki SUZUKIToshio SHIMAMORIAyano SATOKenji TUKANOMasakazu TSUCHIYAJeffrey LAKRITZ
Author information
JOURNAL FREE ACCESS Advance online publication

Article ID: 14-0545

Details
Abstract
The aim of the present study was to compare the ability of the commercially available portable test system (PTSTM) to detect endotoxin activity in bovine serum, with that of the traditional LAL-kinetic turbidimetric (KT) and chromogenic (KC) assays. Prior to testing, serum samples, which were obtained from endotoxin-challenged cattle, were diluted 1:20 in endotoxin-free water and heated to 80°C for 10 min. The performance of the PTSTM was not significantly different from that of the traditional LAL-based assays. The results using PTSTM correlated with those using KT (r2=0.963, P<0.001) or KC assays (r2=0.982, P<0.001). Based on these findings, the PTSTM could be applied as a simplified system to assess endotoxin activity in bovine serum.
Content from these authors
© 2015 by the Japanese Society of Veterinary Science

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.ja
feedback
Top