Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
高速走査形質量分析計
近藤 敏郎渡辺 一生
著者情報
ジャーナル フリー

1966 年 14 巻 5 号 p. 257-264

詳細
抄録
The ion detector for a fast scanning mass spectrometer has been developed. The instrument used is a Hitachi mass spectrometer: modified RMU-6D, of a single focusing, 90° sector type. A mass range of m/e 10-500 may be magnetically scanned in 3 seconds: the peaks are recorded on strip chart of are cording galvanometer.
The ion detector is described which incorporates a electron-multiplier and a fast response electrometer. The electron multiplier consists of 10 stage Ag-Mg dynodes. The electrometer has the feedback resistor sorunded a metal sleeve which is connected the output of the amplifier through a P.I.D. network and the response speed is increased.
The response and the noise of the ion detector are discussed. Design procedures led from the discussion and the complete circuits are given. The ion detector can be used in measurement of current from 2.5×10-11 to 1×10-14 amperes and the 90% inditial response is 40 microseconds.
著者関連情報
© 日本質量分析学会
前の記事 次の記事
feedback
Top