長波長放射計の検定について(I)-放射収支計の場合
嘉納 宗靖鈴木 正矢田 明
著者情報
ジャーナル フリー

1973 年 24 巻 2 号 p. 249-261

詳細
抄録
The effects of the environmental wind and temperature fields on the measurement of net flux of radiation by conventional net radiometers are examined. The variation in output of the net radiometer due to the wind field exceeds ten percent for both FUNK'S net radiometer ("wind shield-type" net radiometer) and GIER and DUNKLE'S net radiometer ("ventilation-type" net radiometer) in the ranges of wind speeds, vertical temperature gradients and incident radiation fluxes concerned. And as for FUNK'S net radiometer the more accurate value of net flux of radiation is expected to be obtained in the windy condition rather than in the calm condition.
The devices hitherto available f or calibrating net radiometers for longwave radiation are discussed in brief and our new device is presented.