Demonstration of Human Fetal Bone Morphology with MR Imaging: A Preliminary Study
Yoshiko MatsubaraToru HigakiChihiro TaniShogo KamiokaKuniaki HaradaHirohiko AoyamaYuko NakamuraTomoyuki AkitaKazuo Awai
著者情報
ジャーナル オープンアクセス
電子付録

2020 年 19 巻 4 号 p. 310-317

詳細
抄録

Purpose: CT is a useful modality for the evaluation of fetal skeletal dysplasia but radiation exposure is unavoidable. The purpose of this study is to compare the usefulness of MRI and CT for evaluating the fetal skeletal shape.

Methods: This study was approved by our Institutional Review Board. Fetal specimens (n = 14) were scanned on a 3T MRI scanner using our newly-developed sequence. It is based on T2*-weighted imaging (TR, 12 ms; TE for opposed-phase imaging, 6.1 ms, for in-phase imaging, 7.3 ms; flip angle, 40°). The specimens were also scanned on a 320 detector-row CT scanner. Four radiologists visually graded and compared the visibility of the bone shape of eight regions on MRI- and CT-scans using a 5-point grading system.

Results: The diagnostic ability of MRI with respect to the 5th metacarpals, femur, fibula, and pelvis was superior to CT (all, P < 0.050); there was no significant difference in the evaluation results of observers with respect to the cervical and lumbar spine, and the 5th metatarsal (0.058 ≤ P ≤ 1.000). However, the diagnostic ability of MRI was significantly inferior to CT for the assessment of the bone shape of the thoracic spine (observers A and C: P = 0.002, observers B and D: P = 0.001).

Conclusion: The MRI method we developed represents a potential alternative to CT imaging for the evaluation of the fetal bone structure.