Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Major Papers
Visualization of Coronary Arterial Wall Based on Maximum Intensity Fusion of Whole-Heart MR Angiograms and Water Suppression SPIR 3D T1 TFE Images
Shigeko TANAKA, Masuki MORI, Kazuhisa KITAZAKI, Koichi NAKAHIRA, Hiroki NAKATSUKA, Yasutaka MAEDA, Hisashi MADONO, Yukari NAITO, Yuko SUGIMORI, Yuichi INOUE
著者情報
ジャーナル オープンアクセス

2009 年 8 巻 2 号 p. 55-63

詳細
抄録
Purpose: We estimated the coronary artery wall using maximum intensity fusion (MIF) of whole-heart magnetic resonance (MR) angiography (WHCA) and water suppression-spectral presaturation with inversion recovery (WS-SPIR) 3D T1-weighted turbo field echo (3DT1 TFE).
Methods: We created a phantom using a wall of plastic bottles varied with plastic tapes measuring 0.4 to 3.0 mm thick (0-14 sheets) by vernier caliper and compared widths with those on profile curves.
In 3 patients, to clarify the capacity to visualize the coronary wall in vulnerable plaque, we acquired WS-SPIR 3D T1 TFE and WS-spectral attenuation with inversion recovery (SPAIR) (inversion time [TI] 400 ms) 3D T1 TFE images of carotid vulnerable plaque; also termed “lipid-rich plaque,” vulnerable plaque is considered to be visualized in high intensity.
We utilized the same geometric parameters and rest period on WHCA as for WS-SPIR 3D T1 TFE. We obtained MIF of WHCA and WS-SPIR 3D T1 TFE and measured thickness of the right coronary artery (RCA) wall on the profile curve in 18 cases.
Results: The widths of the dip of the lower third of the bottom to head on the profile curve were consistent with actual measurement at 1-2 mm, the usual coronary artery wall thickness. Carotid plaques of high intensity by T1-weighted black-blood (T1BB) and T2-weighted BB (T2BB) methods showed high intensity on WS-SPAIR (TI 400 ms) 3D T1 TFE and low intensity on WS-SPIR 3D T1 TFE. With or without vulnerable plaque in the coronary artery wall, MIF of WHCA and WS-SPIR 3D T1 TFE reflected the coronary artery wall. We obtained bands of low intensity in MIF between epicardial fat of WS-SPIR 3D T1 TFE and coronary artery lumen of WHCA all but mid RCA in all 18 cases. We were unable to detect mid RCA in 5 cases. The outline of the obstructed mid RCA in 1 case was clear in WS-SPIR 3D T1 TFE. The higher velocity of RCA movement caused blurring in another 4 cases in both WHCA and WS-SPIR 3D T1 TFE. Those wall thickness of proximal or mid RCA averaged 1.3±0.2 mm.
Conclusion: Bands of low intensity between epicardial fat and coronary artery lumen on MIF of WHCA and WS-SPIR 3D T1 TFE can reflect the coronary artery wall.
著者関連情報
© 2009 by Japanese Society for Magnetic Resonance in Medicine
前の記事 次の記事
feedback
Top