Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Original Articles
Novel Intraoperative Cerebral Blood Flow Monitoring by Laser-Doppler Scanner
Hiroyuki NAKASETakanobu KAIDOShuzo OKUNOTohru HOSHIDAToshisuke SAKAKI
著者情報
ジャーナル オープンアクセス

2002 年 42 巻 1 号 p. 1-4

詳細
抄録
Laser-Doppler (LD) flowmetry was used to measure tissue perfusion non-invasively and continuously during neurosurgical operations using an LD scanner. Scanning was usually completed in 20 seconds. Measurements were processed in software to provide a color-coded image of the tissue perfusion. Moreover, the measurement data, expressed in LD-units, could be used for statistical data analysis. No physical contact was necessary between the scanning device and the exposed brain tissue. The imager provided two-dimensional microvascular flow maps non-invasively and quantitatively during brain operations, and could show the CO2 reactivity in the vessels. LD scanning flowmetry is a promising intraoperative monitoring method for cerebral blood flow changes.
著者関連情報
© 2002 by The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/