Shortening of Wallstent RP During Carotid Artery Stenting Requires Appropriate Stent Placement
Hiroshi AIKAWAShun-ichi NAGATAMasanari ONIZUKAMasanori TSUTSUMIMinoru IKOTomonobu KODAMAKouhei NIIShuko MATSUBARAHosei ETOUYoshinori GOKiyoshi KAZEKAWA
著者情報
ジャーナル オープンアクセス

2008 年 48 巻 6 号 p. 249-253

詳細
抄録

Changes in the location and length of the Wallstent RP during carotid artery stenting (CAS) were evaluated using intraoperative videos of 28 patients with carotid artery stenosis who underwent CAS with a 10/20 mm Wallstent RP to determine the appropriate stent placement. The stent was deployed after its midpoint was positioned over a virtual center line, the perpendicular line which crossed the most stenotic point of the lesion on the road mapping image. The length of the stenotic lesion, the changes in the locations of the distal and proximal ends of the stent, and the changes in stent length were examined. The distal end of the stent moved a maximum of 6.1 mm toward the proximal side to a point 19.9 mm from the virtual center line. The proximal end moved a maximum of 11.3 mm toward the distal side to a point 14.7 mm from the virtual center line. The stent length ranged from 37.7 to 44.5 mm (mean 41.2 mm). The 10/20 mm Wallstent RP placed by our technique covers the entire lesion with no less than 5.7 mm of margin over the segment distal to the lesion in patients with stenotic segments shorter than 29.4 mm.

著者関連情報
© 2008 by The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
前の記事 次の記事
feedback
Top