Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Technical Note
New Titanium Spacer for Cervical Laminoplasty: Initial Clinical Experience
—Technical Note—
Satoshi TANI, Futoshi SUETSUA, Junichi MIZUNO, Hisaaki UCHIKADO, Hiroyasu NAGASHIMA, Masahiko AKIYAMA, Akira ISOSHIMA, Hiroki OHASHI, Yoshitaka HIRANO, Toshiaki ABE
著者情報
ジャーナル オープンアクセス

2010 年 50 巻 12 号 p. 1132-1136

詳細
抄録
Many commercially available hydroxyapatite (HA) spacers for cervical laminoplasty have been introduced but have disadvantages such as lack of plasticity, easy cracking, and occasional difficulty in fixation by sutures. Here we present the short-term results of a newly designed titanium spacer (Laminoplasty Basket) in open-door cervical laminoplasty, and evaluated clinically and radiologically. The titanium box-shaped spacer with two arms for fixation was easily inserted and fixed into the laminoplasty space with 4-mm or 5-mm length screws after the posterior cervical arch was repositioned for the canal expansion. Twenty-one patients with cervical myelopathy due to spondylosis or ossification of the longitudinal ligament or developmental narrow canal observed for more than 6 months postoperatively were enrolled in this study. The neurological condition of these patients improved from 9.4 points on the Japanese Orthopaedic Association scale preoperatively to 13.5 points at 6 months after surgery. Postoperative radiological evaluation showed no laminar closure or implant failure and cervical spine curvature was maintained. These results seemed to have no significant difference compared with those using HA spacers. This titanium spacer is a potential substitute for conventional HA or other similar devices in cervical laminoplasty.
著者関連情報
© 2010 by The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
前の記事 次の記事
feedback
Top