Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Technical Note
New Instrument for Creating Bone Graft Cavities for Anterior Cervical Decompression and Fusion: the Anterior Fusion Spinal Fork
—Technical Note—
Tomosato YAMAZAKISusumu YASUDAKazuya UEMURAAkira MATSUMURA
著者情報
ジャーナル オープンアクセス

2012 年 52 巻 9 号 p. 691-694

詳細
抄録
Sufficient bone decompression of osteophytes is important for positive functional outcomes in anterior cervical spine surgery. Achieving good alignment and bone fusion in anterior cervical decompression and fusion requires a bone graft bed of the optimum size and shape. We have developed a stainless steel instrument named the anterior fusion spinal fork, which is designed to aid in accurately drilling the bone cavity, thus enabling selection of the correct size of bone graft or bone graft substitutes. The device has an open design with 4 prongs, and resembles a three-dimensional fork. This instrument assists in guiding the direction of drilling, and marking the drilling point of the graft cavity with pyoctanin markers. We have used this instrument in 40 cases of anterior cervical spine surgery using the modified Smith-Robinson procedure since March 2000. This simple instrument allowed us to design and construct a bone graft cavity of the correct size and shape for iliac bone graft or bone graft substitutes for anterior cervical spine surgery.
著者関連情報
© 2012 by The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.