Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Technical Notes
Usefulness of a New Electronic Conductivity Device with a Pedicle Probe and a Multi-axis Angiography Unit for Inserting a C1 Lateral Mass Screw Safely and Tightly: A Technical Note
Hiroto KAGEYAMAShinichi YOSHIMURAKenichi MATSUDAYasunori YOSHIDAHidetoshi MATSUKAWAKiyofumi YAMADA
Author information
JOURNAL OPEN ACCESS

2019 Volume 59 Issue 12 Pages 523-528

Details
Abstract

The C1 lateral mass screw (LMS) is widely used as one of the screws for atlantoaxial fixation. Tight bicortical screwing from the posterior to anterior cortical margin of the atlas is recommended. However, important structures, such as the internal carotid artery, are located around this area so precision is required to avoid injuring them. We describe the usefulness of a new electronic conductivity device (ECD) with a pedicle probe and a multi-axis angiography unit for inserting the C1 LMS. Four consecutive patients who were treated with C1 and C2 posterior fixation using an ECD and a multi-axis angiography unit in the hybrid operating room were included. All patients were treated successfully. Seven of eight bicortical screws could be inserted into the perfectly ideal location. The median (interquartile range) distance from the anterior margin of the atlas to the tip of the screw was 0.81 mm (0.43, 1.21 mm). This study suggested that the ECD and multi-axis angiography unit are useful for inserting the C1 LMS safely and tightly.

Content from these authors
© 2019 by The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
Previous article Next article
feedback
Top