Primary Experiences with Robot-assisted Navigation-based Frameless Stereo-electroencephalography: Higher Accuracy than Neuronavigation-guided Manual Adjustment
Yuichiro KOJIMATakehiro UDAToshiyuki KAWASHIMASaya KOHMasato HATTORIYuki MITONoritsugu KUNIHIROShohei IKEDARyoko UMABATakeo GOTO
著者情報
ジャーナル オープンアクセス

2022 年 62 巻 8 号 p. 361-368

詳細
抄録

The use of robot-assisted frameless stereotactic electroencephalography (SEEG) is becoming more common. Among available robotic arms, Stealth Autoguide (SA) (Medtronic, Minneapolis, MN, USA) functions as an optional instrument of the neuronavigation system. The aims of this study were to present our primary experiences with SEEG using SA and to compare the accuracy of implantation between SA and navigation-guided manual adjustment (MA). Seventeen electrodes from two patients who underwent SEEG with SA and 18 electrodes from four patients with MA were retrospectively reviewed. We measured the distance between the planned location and the actual location at entry (De) and the target (Dt) in each electrode. The length of the trajectory did not show a strong correlation with Dt in SA (Pearson's correlation coefficient [r] = 0.099, p = 0.706) or MA (r = 0.233, p = 0.351). De and Dt in SA were shorter than those in MA (1.99 ± 0.90 vs 4.29 ± 1.92 mm, p = 0.0002; 3.59 ± 2.22 vs 5.12 ± 1.40 mm, p = 0.0065, respectively). SA offered higher accuracy than MA both at entry and target. Surgical times per electrode were 38.9 and 32 min in the two patients with SA and ranged from 51.6 to 88.5 min in the four patients with MA. During the implantation period of 10.3 ± 3.6 days, no patients experienced any complications.

著者関連情報
© 2022 The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
前の記事 次の記事
feedback
Top