抄録
A novel x-ray tube with a line filament has been developed for the in-situ calibration of the x-ray crystal spectrometer (XCS) in the KSTAR machine. The characteristics of the x-ray tube are investigated from the x-ray images obtained by using a pinhole and a CCD detector. It is found that the image has the width of about 0.1 mm, which is much improved as compared with the previous experimental results. In addition, there is a uniform region around the center of the image within its full length of 13.5 mm. This work may lead to the development of a novel x-ray tube with a line focus, which is required for the calibration of the XCS. Experimental results from the investigation of the x-ray tube are presented and the technical issues in a design of the in-situ calibration system using the x-ray tube for the KSTAR XCS are discussed.