脳卒中の外科
Online ISSN : 1880-4683
Print ISSN : 0914-5508
ISSN-L : 0914-5508
原  著
Double balloon protectionを用いた頚動脈ステント留置術でのintoleranceの検討
綾部 純一久保 篤彦渡辺 正英前田 昌宏田中 悠介田中 良英
著者情報
ジャーナル フリー

2013 年 41 巻 6 号 p. 435-439

詳細
抄録
Balloon protection effectively prevents distal embolism during carotid artery stenting (CAS). However, some patients have occlusion intolerance to transient balloon-induced ischemia and can develop secondary perioperative ischemic complications. We conducted this study to identify factors that can predict ischemic intolerance. We analyzed 58 patients undergoing CAS in which double balloon protection was used. Seven patients (12.1%) had ischemic intolerance. These patients had a significantly lower stenosis rate (p=0.0255), poor collateral flow (p=0.0001), and pooling appearance of the contrast medium in the ipsilateral internal carotid artery during test proximal balloon occlusion (P=0.0086).
We conclude that preoperative assessment of collateral flow and intraoperative assessment of flow pattern during the proximal test occlusion are useful predictors of ischemic intolerance. These findings may be useful when choosing the protection devices during CAS or determining whether a shunt system should be used during carotid endarterectomy.
著者関連情報
© 2013 一般社団法人 日本脳卒中の外科学会
前の記事 次の記事
feedback
Top