Intercomparison of Radiation Temperature Scales between PTB and NMIJ from 1000-degree Celsius to 2500-degree Celsius
Fumihiro SakumaLaina MaJuergen Hartmann
著者情報
会議録・要旨集 フリー

p. 13

詳細
抄録
The intercomparison of radiation temperature scales between PTB and NMIJ was carried out in September 2001. PTB sent their radiation thermometer LP3 to Japan and NMIJ made the comparison between the LP3 and NMIJ’s Topcon radiation thermometers with a comparison blackbody from 1000 °C to 2500 °C. Both scales agreed within ±0.8 °C in the whole range.