1985 年 1 巻 4 号 p. 187-191
Surface overheating is a most frequent limiting factor in clinical application of RF capacitive heatig of deep seated tumors. The surface hot spots often occured around the margin of electrodes and were not prevented merely by perfusing cold water into the conventional pads attached to the electrodes. We have designed a new type coolig pad, 33 × 37 cm in size and 2 cm thick, which is put beneath the conventional pad and widely covers the skin surface. In the phantom experiments and clinical trials of deep hyperthermia using Thermotron RF8, the excessive surface heating around the margin of the electrodes completely disappeared by this new cooling pad.