The Tohoku Journal of Experimental Medicine
Online ISSN : 1349-3329
Print ISSN : 0040-8727
ISSN-L : 0040-8727
Expression Forms and Characteristics of Carr-Purcell-Meiboom-Gill Sequence Magnetic Resonance Images
SUSUMU YAMADATAIJI MATSUZAWAKENJI YAMADASHUICHI ONOTAKASHI HISHINUMA
Author information
JOURNAL FREE ACCESS

1989 Volume 157 Issue 3 Pages 185-192

Details
Abstract
YAMADA, S., MATSUZAWA, T., YAMADA, K., ONO, S. and HISHINUMA, T. Expression Forms and Characteristics of Carr-Purcell-Meiboom-Gill Sequence Magnetic Resonance Images. Tohoku J. Exp. Med., 1989, 157 (3), 185-192 -We propose two expression forms of Carr-Purcell-Meiboom-Gill (CPMG) sequence magnetic resonance images using the imaging parameters of recovery time (Tr), echo interval (Te), echo number (n) from 1 to N (number of echoes). An echo image can be expressed as a form of CPMG (Tr, Texn) and the other image synthesized by summing consecutive echo images numbered from n1 to n2 (1≤n1< n2≤N) as a form of CPMG (Tr, Te×n1-n2). The CPMG (Tr, Te×n) images are weighted by T1, T2 or proton density and show the better morphological representation and contrast resolution of brain tumor lesion. The difference of signal intensities among the tissues is enhanced on the synthesized CPMG (Tr, Te×n1-n2) images with the increased signal-to-noise ratio. The calculated T2 images are also observed. These forms expound the scanning conditions and image characteristics and make easy to be compared with the spin echo images.
Content from these authors
© Tohoku University Medical Press
Next article
feedback
Top