A Modified Signal Intensity Equation of Carr-Purcell-Meiboom-Gill Pulse Sequence for MR Imaging
SUSUMU YAMADATAIJU MATSUZAWAKENJI YAMADASEIRO YOSHIOKASHUICHI ONOTAKASHI HISHINUMA
Author information
JOURNAL FREE ACCESS

1989 Volume 158 Issue 3 Pages 203-209

Details
Abstract

YAMADA, S., MATSUZAWA, T., YAMADA, K., YOSHIOKA, S., ONO, S. and HISHINUMA, T. A Modified Signal Intensity Equation of Carr-Purcell-Meiboom- Gill Pulse Sequence for MR Imaging. Tohoku J. Exp. Med., 1989, 158 (3), 203-209-The signal intensity equation of Carr-Purcell-Meiboom-Gill (CPMG) pulse sequence for magnetic resonance imaging was modified to: S(n)=k•M0•exp (-n•Te/T2) [1-exp(-Tr/T1)] where S(n) is a signal intensity of “n-th” echo; k, constant; M0, initial longitudinal magnetization; n, echo number; Te, echo interval; and Tr, recovery time [Tr=TR-(N-1/2)•Te; TR, repetition time; N, number of echoes]. To evaluate the accuracies of T1 and T2 values calculated from this modified equation, a phantom experiment using five tubes filled with 0.5 to 8mM copper sulfate solutions was performed using a 0.14-T resistive whole-bo dy MR scanner and a spectrometer system. The differences between the image values by this equation and the bulk values by the spectrometer were less than 6.2% (mean±S.D., 3.3±2.1%) in T2 and 15.6% (10.4±4.9%) in T1 (except 8mM). By this modification, not only the image T2 value with a high accuracy but the image T1 value can be obtained simultaneously.