Adjustment of Axial Transverse Tomograph
Tadayoshi Matsuda
著者情報
ジャーナル フリー

1968 年 95 巻 4 号 p. 331-336

詳細
抄録
In order to obtain sharply imaged axial transverse tomograms the focal spot of the x-ray tube, rotation center of the unit and that of the film must be arranged on a straight line. The device to meet the requirement is discussed from the theoretical viewpoint and assessed experimentally. Eventual lack of sharpness in the image is corrected by use of a unit with perfectly stable rotation axis.