Spectrophotometer Calibration by a Double Integrating Sphere Reference Light Source and Display Panel Measurement Using Dark Sphere
Tatsuhiko MATSUMOTOShigeo KUBOTATsutomu SHIMURAShuichi HAGATakehiro NAKATSUEJunichi OHSAKO
Author information
JOURNAL RESTRICTED ACCESS

2010 Volume E93.C Issue 11 Pages 1590-1594

Details
Abstract
We succeeded to develop a reference light source in the range of very low luminance using a double integrating sphere system, and calibrated a commercial spectrophotometer below 1 × 10-5cd/m2 levels, which is 1/100 lower than the specified limit for measurement. And we improved measurements in the ultra low luminance range of displays using the calibrated commercial spectrophotometer and a dark sphere to suppress the influence of the surround.