Coupler Design and Analysis of Capacitive Wireless Power Charging for Implantable Medical Devices
Marimo MATSUMOTOMasaya TAMURA
著者情報
ジャーナル 認証あり 早期公開

論文ID: 2021ECP5058

この記事には本公開記事があります。
詳細
抄録

Couplers in a film-type capacitive wireless power charging (CWC) system for an implantable medical device were designed and analyzed in this work. Due to the high conductivity of the human body, two paths contribute to the power transmission, namely a high-frequency current and an electric field. This was confirmed by an equivalent circuit of the system. During analysis of the system, we used pig skin with subcutaneous fat, which has a high affinity with the human body, to search for a highly efficient electrode shape. Subsequently, we fabricated the designed coupler and measured ηmax. An ηmax of 56.6% was obtained for a half-circular coupler with a radius of 20 mm and a distance of 10 mm between adjacent couplers. This study will contribute to the realization of implantable devices that can be recharged during breaks or while sleeping at home, and is expected to significantly reduce the burden on patients.