体外循環中の「オンライン」かつ「非侵襲的」な血中ヘパリン量モニタリングを目指した緩和時間分布解析
上野 総一朗川嶋 大介松浦 功泰田中 綾武居 昌宏
著者情報
ジャーナル オープンアクセス

2025 年 91 巻 942 号 p. 24-00208

詳細
抄録

The relaxation strength Δεm (m: dielectric relaxation number) regarding heparin volume in blood has been extracted by relaxation time distribution (RTD) analysis implementing in electrical impedance spectroscopy (EIS) for the online and non-invasive monitoring of heparin volume in blood extracorporeal circulation. Extracorporeal circulation experiment was performed under conditions where heparin volume in blood varied over time due to metabolism and heparin administration. By applying the RTD analysis with m as an unknown parameter, the Δεm were extracted from dielectric spectra of circulating blood which were measured by an impedance sensor located at a local section in extracorporeal circuit. As a result, four Δεm (m=1, 2, 3 and 4) were extracted by RTD analysis. Among the extracted Δεm, the variation of Δε3 was in good agreement with heparin volume in blood which was indirectly estimated from the activated clotting time (ACT). By comparing the variations of Δε3 and the variation of ACT, the Δε3 reflects the dielectric relaxation due to heparin-induced aggregation of blood cells and heparin complex with other plasma macromolecules. Therefore, the Δε3 has been shown to be one of the most important measurement indexes for the online and non-invasive monitoring of heparin volume in blood extracorporeal circulation.

著者関連情報
© 2025 一般社団法人日本機械学会

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.ja