2005 年 125 巻 3 号 p. 323-325
A method has been proposed for determining the detection limit (LD) from the slope of a semilogarithmic plot of a B/B0 curve in competitive enzyme linked immunosorbent assay (ELISA). As an application, this paper describes a graphic determination of LD from analogue data in the literature. The LD obtained corresponds to the concentration at which the relative standard deviation of concentration estimates is 30%.