ノート
1H核定量核磁気共鳴分光法の標準化のためのラウンドロビン試験による評価
三浦 亨杉本 直樹渡邊 龍一末松 孝子高柳 庸一郎伊藤 喜之斎藤 直樹澤 竜一加藤 毅藤峰 慶徳小池 亮大福 裕子山田 裕子内海 博明鈴木 敏之
著者情報
ジャーナル フリー

2017 年 137 巻 12 号 p. 1543-1553

詳細
抄録

1H quantitative NMR (1H qNMR) is known as a powerful tool for determination of analytes without the need for their identical standards, which is eligible to a primary rate method. 1H qNMR has been already stipulated to an assay for purity determination in Japanese Pharmacopoeia (JP), and then this technique has been also applied in several fields such as pharmaceutical and food sciences. However, there is little information about the accuracy of 1H qNMR so that the further applications into other fields such as industrial chemistry could be constricted. In this study, in order to assess the reliability of 1H qNMR, we designed the round-robin test of 1H qNMR under the basis of the measurement conditions described in JP. 1,4-Bis(trimethylsilyl)benzene-d4 [1,4-BTMSB-d4, 99.9±0.6% (w/w)] and 3,5-bis(trifluoromethyl)benzoic acid [3,5-BTMFBA, 99.96±0.06% (w/w)], which are certified reference materials (CRMs), were adopted to analyte and qNMR reference standard respectively for the accurate evaluation in this test. Six NMR instruments in 5 institutions optimized to 1H qNMR conditions provided the purity 1,4-BTMSB-d4 within acceptable error range. This result represented that 1H qNMR has the capability to determine precisely the value of analyte in practical analytical field and to be set as official analytical method for purity determination or assay of concentration of organic compounds.

著者関連情報
© 2017 The Pharmaceutical Society of Japan