A PediaSat
TM oximetry catheter (PediaSat: Edwards Lifesciences Co., Ltd., Irvine, CA, U. S. A.), which facilitates continuous measurement of central venous oxygen saturation (ScvO
2), may be useful for surgery for pediatric congenital heart disease. We used PediaSat during a bidirectional Glenn shunt. The patient was a 13-month-old boy. Under a diagnosis of left single ventricle (pulmonary atresia, right ventricular hypoplasia, atrial septal defect) and residual left aortic arch/left superior vena cava, a modified right Blalock-Taussig shunt was performed. Cyanosis deteriorated, so a bidirectional Glenn shunt was scheduled. After anesthesia induction, a 4.5 Fr double-lumen (8 cm) PediaSat was inserted through the right internal jugular vein for continuous ScvO
2 monitoring. Furthermore, the probe of a near-infrared, mixed blood oxygen saturation-measuring monitor was attached to the forehead for continuous monitoring of the regional brain tissue mixed blood oxygen saturation (rSO
2) (INVOS
TM 5100C, Covidien; Boulder, CO, U. S. A.). Blockage of the right pulmonary artery and right superior vena cava decreased the oxygen saturation, ScvO
2, and rSO
2, but increased the central venous pressure. Although changes in ScvO
2 were parallel to those in rSO
2, the former showed more marked changes. A combination of ScvO
2 and rSO
2 for monitoring during Glenn shunt may be safer. J. Med. Invest. 60: 272-275, August, 2013
抄録全体を表示