Heavy ion therapy using the energetic
12C beam is successfully under way at HIMAC, Japan. The method is more advantageous than traditional radiation therapy in dose concentration owing to the Bragg peak and high relative biological effectiveness. A research study using the
11C beam for heavy ion therapy in the future has been carried out in order to develop the capability of monitoring the dose distribution. Our group has examined the total energy absorption spectrum of the
11C beam in a plastic scintillator. We could clearly observe the total absorption peak of
11C in the energy spectrum and, in addition, we found a broad bump structure was associated with the peak. The bump area occupies 37% of the total spectrum and it probably affects the dose calculation for an accurate treatment planning. We elucidated the mechanism that leads to the structure of the total energy absorption spectra given by
11C and
12C in a block of plastic scintillator. This paper describes the method in detail and gives experimental analysis results which deal with the bump structure. We could explain the bump structure using the energy spectra caused by the fragmentation reactions.