Superheated Water Chromatography of Low Molecular Weight Polyethylene Glycols with Ultraviolet Detection
Takashi YARITARyoji NAKAJIMAKayori SHIMADAShinichi KINUGASAMasami SHIBUKAWA
著者情報
ジャーナル フリー

2005 年 21 巻 8 号 p. 1001-1003

詳細
抄録
Superheated water chromatography (SWC) with ultraviolet detection was applied to the separation of low molecular weight polyethylene glycols (PEGs). PEG oligomers could be detected sensitively when the detection wavelength was set at 190 nm. The effect of column temperature on the separation of PEG oligomers was investigated. The elution time of all PEG oligomers decreased with increase in the column temperature; linear relationships were obtained between ln k and 1/T. A temperature-programmed SWC separation enabled the baseline separation of a PEG 200 sample within 50 min.
著者関連情報
© 2005 by The Japan Society for Analytical Chemistry