Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Simple Spectrophotometric Flow Injection System with an In-valve Minicolumn for Enhancement during the Determination of Chromium(III) Using EDTA
Somchai LAPANANTNOPPAKHUNSupharphorn KASUWASLucksagoon GANRANOOJaroon JAKMUNEEKate GRUDPAN
著者情報
ジャーナル フリー

2006 年 22 巻 1 号 p. 153-155

詳細
抄録
A simple spectrophotometric flow injection (FI) procedure for the determination of Cr(III) using ethylenediaminetetraacetic acid (EDTA) was developed. An FI system with a column packed with Amberlite IR-120(H) was employed for sample pretreatment. This leads to the possibility of a single standard calibration. A linear calibration in a range of 10 - 27 µg Cr(III) was obtained with a detection limit of 1 µg Cr(III) and RSD of 2% (18 µg Cr(III), n = 12). The proposed procedure was applied for determination of Cr(III) in leachate and dietary supplement samples. The results agreed with those obtained by the standard methods.
著者関連情報
© 2006 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top